Appendix 2


Recommended Instrument Sterilants

AGENTS


EXAMPLES


COMMENTS


Physical:
Steam sterilization
(moist heat)


Autoclave


Effectiveness dependent upon temperature, pressure, and time (e.g. 121°C for 15 min. vs 131°C for 3 min.).
Dry heat
Hot bead sterilizer
Dry chamber
Fast. Instruments must be cooled before contacting tissue.
Ionizing radiation
Gamma radiation
Requires special equipment.
Chemical:
Gas sterilization
Ethylene oxide
Requires 30% or greater relative humidity for effectiveness against spores. Gas is irritating to tissue; all materials require safe airing time. Carcinogenic.
Hydrogen peroxide
(Sterad®)
Not useful for "delicate" items.
Chlorine*
Chlorine dioxide (Clidox® , Alcide®)
A minimum of 6 hours required for sterilization. Presence of organic matter reduces activity. Must be freshly made (<14 days).
Aldehydes*

Formaldehyde (6% sol.)

Glutaraldehyde


For all aldehydes: many hours required for sterilization. Corrosive and irritating. Consult safety representative on proper use. Glutaraldehyde is less irritating and less corrosive than formaldehyde.

*Instruments must be rinsed thoroughly with sterile water or saline to remove chemical sterilants before being used.